Accuracy and reliability of linear measurements using 3-dimensional computed tomographic imaging software for Le Fort I Osteotomy
چکیده انگلیسی

Our purpose was to compare the accuracy and reliability of linear measurements for Le Fort I osteotomy using volume rendering software. We studied 11 dried skulls and used cone-beam computed tomography (CT) to generate 3-dimensional images. Linear measurements were based on craniometric anatomical landmarks that were predefined as specifically used for Le Fort I osteotomy, and identified twice each by 2 radiologists, independently, using Dolphin imaging version 11.5.04.35. A third examiner then made physical measurements using digital calipers. There was a significant difference between Dolphin imaging and the gold standard, particularly in the pterygoid process. The largest difference was 1.85 mm (LLpPtg L). The mean differences between the physical and the 3-dimensional linear measurements ranged from −0.01 to 1.12 mm for examiner 1, and 0 to 1.85 mm for examiner 2. Interexaminer analysis ranged from 0.51 to 0.93. Intraexaminer correlation coefficients ranged from 0.81 to 0.96 and 0.57 to 0.92, for examiners 1 and 2, respectively. We conclude that the Dolphin imaging should be used sparingly during Le Fort I osteotomy.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: British Journal of Oral and Maxillofacial Surgery - Volume 52, Issue 3, March 2014, Pages 258–263
نویسندگان
, , , , , ,