کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3132583 1584152 2013 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Is it possible to improve the accuracy of implants inserted with a stereolithographic surgical guide by reducing the tolerance between mechanical components?
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Is it possible to improve the accuracy of implants inserted with a stereolithographic surgical guide by reducing the tolerance between mechanical components?
چکیده انگلیسی

When using a stereolithographic surgical guide, a potentially clinically relevant error may be the mechanical error caused by the bur guide cylinder gap due to the presence of a rotational allowance of drills in the tubes. The aim of the present study was to determine if it is possible to reduce the total error by limiting the tolerance among the mechanical components and to evaluate its clinical incidence. Sixty implants were inserted in eight totally edentate subjects using the External Hex Safe® system with mechanical components modified to minimize the tolerance. Pre- and postoperative computed tomography images were compared, and the angular deviation was calculated between the planned and the placed implants. The mean angular deviation was 2.02° (range 0.81–3.48°; standard deviation 0.87). The results of the present study show that by limiting the error that originates from mechanical components, the total error could be significantly reduced.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: International Journal of Oral and Maxillofacial Surgery - Volume 42, Issue 7, July 2013, Pages 887–890
نویسندگان
, , , , ,