• Fatigue life for contemporary rotary instruments was tested at room and body temperature.
• Under clinical conditions, fatigue resistance appears to be substantially lower than previously thought.
• Transitional temperatures between nickel-titanium crystal structures were documented and may explain why the change from 20°C to 37°C has such a substantial effect.
IntroductionThe purpose of this study was to evaluate the effect of 2 different temperatures (20°C and 37°C) on the cyclic fatigue life of rotary instruments and correlate the results with martensitic transformation temperatures.MethodsContemporary nickel-titanium rotary instruments (n = 20 each and tip size #25, including Hyflex CM [Coltene, Cuyahoga Falls, OH], TRUShape [Dentsply Tulsa Dental Specialties, Tulsa, OK], Vortex Blue [Dentsply Tulsa Dental Specialties], and ProTaper Universal [Dentsply Tulsa Dental Specialties]) were tested for cyclic fatigue at room temperature (20°C ± 1°C) and at body temperature (37°C ± 1°C). Instruments were rotated until fracture occurred in a simulated canal with an angle curvature of about 60° and a radius curvature of 3 mm; the center of the curvature was 4.5 mm from the instrument tip. The number of cycles to fracture was measured. Phase transformation temperatures for 2 instruments of each brand were analyzed by differential scanning calorimetry. Data were analyzed using the t test and 1-way analysis of variance with the significance level set at 0.05.ResultsFor the tested size and at 20°C, Hyflex CM showed the highest resistance to fracture; no significant difference was found between TRUShape and Vortex Blue, whereas ProTaper Universal showed the lowest resistance to fracture. At 37°C, resistance to fatigue fracture was significantly reduced, up to 85%, for the tested instruments (P < .001); at that temperature, Hyflex CM and Vortex Blue had similar and higher fatigue resistance compared with TRUShape and ProTaper Universal.ConclusionsUnder the conditions of this study, using a novel testing design, immersion in water at simulated body temperature was associated with a marked decrease in the fatigue life of all rotary instruments tested.
Journal: Journal of Endodontics - Volume 42, Issue 5, May 2016, Pages 782–787