Evidence for Reduced Fatigue Resistance of Contemporary Rotary Instruments Exposed to Body Temperature
ترجمه فارسی عنوان
شواهدی برای کاهش خستگی از ابزارهای روتاری معاصر در معرض دمای بدن است
کلمات کلیدی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
چکیده انگلیسی


• Fatigue life for contemporary rotary instruments was tested at room and body temperature.
• Under clinical conditions, fatigue resistance appears to be substantially lower than previously thought.
• Transitional temperatures between nickel-titanium crystal structures were documented and may explain why the change from 20°C to 37°C has such a substantial effect.

IntroductionThe purpose of this study was to evaluate the effect of 2 different temperatures (20°C and 37°C) on the cyclic fatigue life of rotary instruments and correlate the results with martensitic transformation temperatures.MethodsContemporary nickel-titanium rotary instruments (n = 20 each and tip size #25, including Hyflex CM [Coltene, Cuyahoga Falls, OH], TRUShape [Dentsply Tulsa Dental Specialties, Tulsa, OK], Vortex Blue [Dentsply Tulsa Dental Specialties], and ProTaper Universal [Dentsply Tulsa Dental Specialties]) were tested for cyclic fatigue at room temperature (20°C ± 1°C) and at body temperature (37°C ± 1°C). Instruments were rotated until fracture occurred in a simulated canal with an angle curvature of about 60° and a radius curvature of 3 mm; the center of the curvature was 4.5 mm from the instrument tip. The number of cycles to fracture was measured. Phase transformation temperatures for 2 instruments of each brand were analyzed by differential scanning calorimetry. Data were analyzed using the t test and 1-way analysis of variance with the significance level set at 0.05.ResultsFor the tested size and at 20°C, Hyflex CM showed the highest resistance to fracture; no significant difference was found between TRUShape and Vortex Blue, whereas ProTaper Universal showed the lowest resistance to fracture. At 37°C, resistance to fatigue fracture was significantly reduced, up to 85%, for the tested instruments (P < .001); at that temperature, Hyflex CM and Vortex Blue had similar and higher fatigue resistance compared with TRUShape and ProTaper Universal.ConclusionsUnder the conditions of this study, using a novel testing design, immersion in water at simulated body temperature was associated with a marked decrease in the fatigue life of all rotary instruments tested.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Endodontics - Volume 42, Issue 5, May 2016, Pages 782–787
نویسندگان
, , , , , ,