کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3149977 1197488 2007 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of Efficacy of a New Custom-Made Pulse Oximeter Dental Probe in Comparison With the Electrical and Thermal Tests for Assessing Pulp Vitality
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Evaluation of Efficacy of a New Custom-Made Pulse Oximeter Dental Probe in Comparison With the Electrical and Thermal Tests for Assessing Pulp Vitality
چکیده انگلیسی

Pulse oximetry is a noninvasive method of measuring vascular health by evaluating oxygen saturation. This study evaluated the efficacy of a new custom-made pulse oximeter dental probe in comparison with the electrical and thermal tests for assessing pulp vitality. Sensitivity, specificity, negative predictive value, and positive predictive value for each test were calculated by comparing the test results with the actual pulpal status, as evaluated by direct visual inspection. The sensitivity of the pulse oximeter was found to be 1.00, as compared to 0.81 with the cold test and 0.71 with the electrical test. The specificity of the pulse oximeter was 0.95, as compared to 0.92 with the cold and electrical pulp tests. Thus, the custom-made pulse oximeter dental probe is an effective, accurate, and objective method of evaluating pulp vitality.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Endodontics - Volume 33, Issue 4, April 2007, Pages 411–414
نویسندگان
, , ,