کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3251954 1207023 2013 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Laser etching causing fatigue fracture at the neck–shoulder junction of an uncemented femoral stem: A case report
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی طب اورژانس
پیش نمایش صفحه اول مقاله
Laser etching causing fatigue fracture at the neck–shoulder junction of an uncemented femoral stem: A case report
چکیده انگلیسی

Fatigue fracture of a femoral component in total hip arthroplasty is a rare occurrence but well documented in the literature. It is understood that proximal loosing of a femoral stem with a well fixed stem distally will result in cantilever bending and eventual fatigue fracture of the stem. Other factors which may potentiate a fatigue fracture are material design, implant positioning, and patient characteristics. More recently, laser etching on the femoral neck of an implant has resulted in fatigue fracture. We report a case of a fatigue fracture at the neck–shoulder junction in a well fixed, uncemented, femoral component due to laser etching in the region of high tensile stress.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Orthopaedics - Volume 10, Issue 2, June 2013, Pages 95–98
نویسندگان
, , , ,