کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3482819 1596834 2007 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Investigation of a new design for zirconia dental implants
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی پزشکی و دندانپزشکی (عمومی)
پیش نمایش صفحه اول مقاله
Investigation of a new design for zirconia dental implants
چکیده انگلیسی

ObjectiveTo assess the biomechanical properties of a new design configuration for zirconia dental implants.MethodsThe new design has a cylindrical shape that is partially hollow and porous in the bottom, which permits the implants to be locked into the alveolar bone over time. It also utilizes bioactive glass coatings to increase adhesion to surrounding bone structure. Samples of the new design were fabricated in the laboratory and their material strength, hardness, and fracture toughness were evaluated. In addition, biocompatibility of the new design was evaluated through testing in dogs.ResultsResults of mechanical tests indicate that structural properties of the new design exceed the usual requirements for implants. Moreover, animal tests suggest that there is appreciable improvement in lock-in strength and osteointegration.ConclusionThe new design configuration is biomechanically feasible and further research is warranted to improve the design for human use.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Medical Colleges of PLA - Volume 22, Issue 5, October 2007, Pages 303-311