کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3866268 1598917 2013 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Patterns of Hardware Related Electrode Failures in Sacral Nerve Stimulation Devices
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیماری‌های کلیوی
پیش نمایش صفحه اول مقاله
Patterns of Hardware Related Electrode Failures in Sacral Nerve Stimulation Devices
چکیده انگلیسی
Abnormal electrical impedance occurred in approximately 13% of cases permanently implanted in our series. Short circuits presented earlier and often required surgical intervention. Open circuits presented later and may have potentially been secondary to microfractures that accumulate with time at the sacral plate, resulting in later presentation. Almost a third of patients with abnormal electrical impedance associated with clinical inefficacy were treated conservatively, primarily with reprogramming.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Urology - Volume 190, Issue 1, July 2013, Pages 175-179