کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3879521 1599021 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Frequency of Ureteroscope Damage Seen at a Tertiary Care Center
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیماری‌های کلیوی
پیش نمایش صفحه اول مقاله
Frequency of Ureteroscope Damage Seen at a Tertiary Care Center
چکیده انگلیسی

PurposeThere is controversy regarding ureteroscope durability. Little is known regarding the subsequent durability of a flexible ureteroscope after major damage has been incurred and the ureteroscope has been repaired. Maintenance and repair are associated with significant cost. We reviewed and assessed the frequency and cause of ureteroscope damage at our medical center.Materials and MethodsFrom December 2001 we prospectively recorded the specific use of all ureteroscopes and any resultant damage at a single tertiary care institution. We then reviewed a total of 601 ureteroscopic cases involving 654 semirigid and flexible ureteroscope uses from December 2001 to November 2004. Cases were performed by multiple residents and fellows under the supervision of 3 attending urologists (CML, RJL and VGB). Retrograde and antegrade cases involving stones, urothelial carcinoma, strictures and diagnostic evaluations were included. Repairs for the respective ureteroscopes were performed by the original manufacturer.ResultsA total of 53 reports of damage (8.1% of total uses) were recorded. Major damage when the scope was deemed unusable and required repair was seen in 39 cases (6.0%). Four newly purchased flexible ureteroscopes were entered into the study and they provided 40 to 48 uses before the initial repair was needed. After these new ureteroscopes underwent comprehensive repair for major damage they averaged only 11.1 uses (median 8) before needing repair again. Older model ureteroscopes that underwent repair before being entered into our study averaged between 4.75 and 7.7 uses before being sent for subsequent repair. Of the total of 39 breakages 39 for which ureteroscopes were sent for repair 14 (35.9%) were the result of errant laser firing, 11 (28.2%) were the result of excessive torque, 8 (20.5% 8) were the result of decreased flexion in the distal tip or another loss of function without obvious iatrogenic cause, 3 (7.7%) were the result of multifocal catastrophic damage involving laser firing and excessive torque, and 3 (7.7%) were the result of cleaning and processing outside of the ureteroscopy suite.ConclusionsThe most important risk factors for predicting the number of uses expected from a ureteroscope at our institution is ureteroscope age and whether the ureteroscope has undergone comprehensive repair as the result of prior damage. Our analysis suggests that after damage occurs to a ureteroscope more damage occurs with greater frequency. The cost of maintaining previously used ureteroscopes should be carefully considered in comparison to the cost of purchasing a new ureteroscope.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Urology - Volume 176, Issue 2, August 2006, Pages 607–610
نویسندگان
, , , , , ,