We tested a new bedside method to determine the function of native arteriovenous fistula in 16 patients performed during hemodialysis without stopping the treatment. We initially measured vascular access flow (Qa) in each patient using the Transonic HD01plus device. We then measured the pressure in arterial and venous drip chambers at different blood pump flow rates (Qbset=0, 50, 100, 250, 300, 350 ml/min). The intravascular blood pressure gradient (Pf) between arterial and venous puncture sites was estimated by a mathematical model. Pf was positive for low Qbset, but became negative when Qbset overcame the threshold value (QInv). Such critical flow showed a high correlation with Qa, even if it was systemically lower. Computer analysis of fluid dynamics showed that when the blood pump flow overcame the QInv threshold, a critical transition from laminar flow to vortex circulation took place downstream of the venous needle, causing a dangerous shearstress on the vessel wall. Our results show that QInv provides an indication of the maximal blood pump flow rate needed to be reached to maximize blood flow supply in order to limit hemodynamic stress on the vascular access.
Journal: Kidney International - Volume 75, Issue 5, 1 March 2009, Pages 550–557