کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
3926202 | 1253144 | 2006 | 7 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: A New Artificial Urinary Sphincter with Conditional Occlusion for Stress Urinary Incontinence: Preliminary Clinical Results A New Artificial Urinary Sphincter with Conditional Occlusion for Stress Urinary Incontinence: Preliminary Clinical Results](/preview/png/3926202.png)
ObjectivesTo perform a preliminary clinical investigation to determine the safety and efficacy of a novel artificial urinary sphincter (AUS) with conditional occlusion for the treatment of stress urinary incontinence.MethodsMale patients with urodynamically proven stress urinary incontinence after a prostatectomy were implanted with the novel AUS. They were followed up over a period of 12 months and the device tested for efficacy by using objective measurements of urinary leakage and continence. We derived a new measure for continence called the Continence Index.ResultsWe have demonstrated that the patients receiving the new AUS showed a reduction of greater than 10-fold in mean daily leakage volume from 770.6 ml to 55.1 ml. There was an overall improvement in the Continence Index from 54% to 97%.ConclusionsThe new AUS with conditional occlusion provides good continence rates and enables adjustment of regulating pressure in situ.
Journal: European Urology - Volume 50, Issue 3, September 2006, Pages 574–580