کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3956020 1255285 2015 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Management of Perforated Essure with Migration into Small and Large Bowel Mesentery
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی زنان، زایمان و بهداشت زنان
پیش نمایش صفحه اول مقاله
Management of Perforated Essure with Migration into Small and Large Bowel Mesentery
چکیده انگلیسی
The Essure permanent birth control system (Conceptus Inc, San Carlos, CA) is currently the only Food and Drug Administration-approved hysteroscopic sterilization method and has been widely accepted as a safe and effective procedure. We present a rare case of tubal perforation, coil fragmentation, and distal migration into small and large bowel mesentery 8 days after the insertion of the Essure device. We describe the successful management of this complication using laparoscopy and intraoperative fluoroscopy. Providers using Essure must be aware of the possibility of fragmentation of the Essure coils. Intraoperative imaging, ideally fluoroscopy, should be strongly considered in the management of Essure migration to ensure localization and full retrieval of Essure material.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Minimally Invasive Gynecology - Volume 22, Issue 3, March–April 2015, Pages 504-508