The Risk of Capsular Breakage from Phacoemulsification Needle Contact With the Lens Capsule: A Laboratory Study
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
The Risk of Capsular Breakage from Phacoemulsification Needle Contact With the Lens Capsule: A Laboratory Study
چکیده انگلیسی
Factors influencing capsular breakage may include active vacuum at the tip, flow rate, needle gauge, and sharpness. Nineteen-gauge sharp tips were more likely than 20-gauge tips to cause breakage in lower vacuum methods. For higher-vacuum methods, breakage is more likely with 20-gauge than with 19-gauge tips. Rounded-edge tips are less likely than sharp-edged tips to cause breakage.