کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4004317 1602208 2008 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A Fluidics Comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature Phacoemulsification Machines
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
A Fluidics Comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature Phacoemulsification Machines
چکیده انگلیسی
Machine-indicated accuracy, POS, and UFV were statistically significantly different. Signature had the lowest POS and vacuum to maintain flow. Regarding POS, Stellaris was close to Signature; regarding vacuum to maintain flow, Infiniti and Stellaris were similar. Minimizing POS and vacuum to maintain flow potentially are important in avoiding ocular damage and surgical complications.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: American Journal of Ophthalmology - Volume 145, Issue 6, June 2008, Pages 1014-1017.e1
نویسندگان
, , , ,