Field evaluation of the Welch Allyn SureSight™ Vision Screener: Incorporating the Vision in Preschoolers study recommendations
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
Field evaluation of the Welch Allyn SureSight™ Vision Screener: Incorporating the Vision in Preschoolers study recommendations
چکیده انگلیسی

IntroductionThe prospective Vision in Preschoolers (VIP) study evaluated 11 methods of screening and proposed referral criteria for the Welch Allyn SureSight™ Vision Screener with 90% and 94% specificity. The SureSight had a higher sensitivity than most other screening techniques when these criteria were applied. We evaluated the usefulness of these criteria in a field study of healthy preschool children.MethodsThe SureSight software was altered to recommend referral using the VIP referral criteria with 90% specificity. Lions Club volunteers screened preschool children throughout Tennessee. Referred children underwent comprehensive eye examinations with cycloplegic refraction. Examination failure criteria were based upon published standards. Reanalysis using the 94% specificity criteria was then performed. Outcomes included referral rate and positive predictive value.ResultsThe SureSight was used to screen 4,733 children, and screening was successful in 99.7% of children. The referral rate using the 90% specificity criteria was 12.2%. Most children (73%) were referred for suspected astigmatism. The positive predictive value was 30%. Using the 94% specificity criteria from the VIP study decreased the referral rate to 7.9% and substantially decreased over referral for suspected astigmatism; however, several anisometropes went undetected. Higher specificity was achieved by raising astigmatism referral criteria to 2.2 diopters while leaving the anisometropia criteria unchanged.ConclusionsThe SureSight can be used successfully for preschool screening in the field provided that criteria with high specificity are incorporated into the instrument’s software program. Higher rates of positive predictive value can be achieved without jeopardizing sensitivity by raising astigmatism referral criteria to 2.2 diopters.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of American Association for Pediatric Ophthalmology and Strabismus - Volume 11, Issue 3, June 2007, Pages 243–248
نویسندگان
, , , , , ,