Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system
چکیده انگلیسی
Ultrasound pachymetry and Artemis 2 CCT measurements were highly correlated; the 11 μm mean difference in measurements may be attributed to decentration, oblique incidence of the probe to the cornea, or possibly the effect of topical anesthesia with contact pachymetry. Although the mean difference between Orbscan II and Artemis 2 values was 7.5 μm, Orbscan values were less correlated than Artemis 2 values with contact US pachymetry and were prone to underestimation of the CCT in thinner corneas.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cataract & Refractive Surgery - Volume 34, Issue 11, November 2008, Pages 1906-1912