Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia
چکیده انگلیسی
Refractive surgery had a modest effect on the agreement between Pentacam readings and US measurements. With Orbscan II, the 95% LoA width nearly doubled after surgery. Although the Pentacam seems to show better agreement than Orbscan II, especially after refractive surgery, it is not advisable to use the 3 devices interchangeably in every clinical situation.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cataract & Refractive Surgery - Volume 33, Issue 10, October 2007, Pages 1701-1707