کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4020960 | 1262047 | 2007 | 5 صفحه PDF | دانلود رایگان |

Endothelial decompensation is a serious complication of phakic intraocular lens (pIOL) implantation and is a major concern during the postoperative period. We report 3 eyes in which the same foldable angle-supported pIOL was implanted to correct high myopia. Rapid and severe postoperative endothelial cell loss occurred in all 3 eyes. An over-sized pIOL that induced excessive vaulting into the anterior chamber was the main risk factor. In 2 eyes, the pIOL was explanted uneventfully; 1 eye required Descemet's stripping automated endothelial keratoplasty because of total endothelial decompensation. These cases illustrate the importance of accurate sizing of foldable angle-supported anterior chamber pIOLs to avoid excessive vaulting. They also highlight the importance of regular follow-up and preventive pIOL explantation as soon as significant endothelial cell loss is detected.
Journal: Journal of Cataract & Refractive Surgery - Volume 33, Issue 8, August 2007, Pages 1477–1481