Reproducibility of Macular Thickness and Volume Using Zeiss Optical Coherence Tomography in Patients with Diabetic Macular Edema
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
Reproducibility of Macular Thickness and Volume Using Zeiss Optical Coherence Tomography in Patients with Diabetic Macular Edema
چکیده انگلیسی
Reproducibility is better for central subfield thickness measurements than for center point measurements, and variability is less with retinal thickness when expressed as a percent change than when expressed as an absolute change. A change in central subfield thickness exceeding 11% is likely to be real. Scans with an SD of the center point of ≥10.0% are less reproducible and should be viewed with caution when assessing the validity of an observed change in retinal thickness in patients with DME.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ophthalmology - Volume 114, Issue 8, August 2007, Pages 1520-1525