Effects of Corneal Thickness, Corneal Curvature, and Intraocular Pressure Level on Goldmann Applanation Tonometry and Dynamic Contour Tonometry
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
پیش نمایش صفحه اول مقاله
Effects of Corneal Thickness, Corneal Curvature, and Intraocular Pressure Level on Goldmann Applanation Tonometry and Dynamic Contour Tonometry
چکیده انگلیسی
Intraocular pressure measured by GAT was consistently lower when compared with DCT, and this difference was greatest with thinner CCT. Dynamic contour tonometry was also less affected by variations in CCT. Corneal curvature affected IOP measurements with DCT but not GAT, but this effect was less than the CCT effect on GAT. Goldmann applanation tonometry tended to underestimate IOP at higher levels and overestimate it at lower IOP levels when compared to DCT.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ophthalmology - Volume 114, Issue 1, January 2007, Pages 20-26