Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: Measurements reproductibility with EOS imaging system versus conventional radiographies
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: Measurements reproductibility with EOS imaging system versus conventional radiographies
چکیده انگلیسی

SummaryIntroductionThe literature has recently underlined the interest of pelvic and acetabular component orientation measurements in the standing and sitting position. Radiographic follow-up of total hip arthroplasty (THA) is based on standard AP and lateral X-rays. The use of EOS™ 2D imaging system reduces patient's radiation exposure compared to conventional X-rays. However, using this system, the validity and reproducibility of angular measurements, have not been studied yet for the measurement of pelvic and acetabular parameters in patients with THA.HypothesisThe EOS™ 2D imaging system offers similar advantages to conventional X-rays in the measurement of pelvic and acetabular orientation parameters which are commonly used.Patients and methodFive angular parameters characterizing pelvic tilt and acetabular cup orientation were determined using the same digital measurement Imagika™ software based on two series of standard X-rays and EOS™ 2D images acquired in both standing and sitting positions. Radiographs from 50 patients with unilateral THA were measured three times by two observers. Intra- and interobserver reproducibility using each method was independently studied then paired comparison was performed.ResultsThe ICC and Spearman rank correlation coefficient demonstrated an excellent EOS/conventional X-ray correlation. According to the parameters, the mean difference between these two imaging modalities ranged from 0.30° to 3.43° (P < 0.05). The intra- and interobserver variability ranged from ± 2.97° to ± 6.46° using the EOS™ imaging system and from ± 4.26° to ± 10.22° using conventional X-rays (P < 0.05).DiscussionThe EOS™ 2D imaging system may replace conventional X-rays in the assessment and monitoring of pelvic and acetabular cup orientation in THA.Level of evidenceLevel III. Prospective diagnostic study.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Orthopaedics & Traumatology: Surgery & Research - Volume 97, Issue 4, June 2011, Pages 373–380
نویسندگان
, , , , , , ,