کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4112729 | 1606039 | 2011 | 4 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: Cochlear implant electrode array misplaced in Hyrtl's fissure Cochlear implant electrode array misplaced in Hyrtl's fissure](/preview/png/4112729.png)
Hyrtl's fissure is a cleft that is present in the developing fetal petrous temporal bone and extends from the area inferior to the round window to the meninges of the posterior fossa. Persistent Hyrtl's fissure, due to incomplete ossification, is considered a rare temporal bone malformation, and is a known cause of perilabyrinthine cerebrospinal fluid fistula.Very few cases are reported as being at risk of complication of cochlear implant surgery. Here we report the case of an 8-year-old boy with misplacement of an electrode array in Hyrtl's fissure. The diagnosis was made postoperatively, since cochlear implant failure was suspected from non-auditory responses. Computed tomography (CT) revealed the extracochlear location of the electrode array. We emphasize the role of presurgical imaging CT and magnetic resonance imaging in detecting temporal bone abnormalities, and we discuss the value of intraoperative auditory nerve response telemetry and postoperative radiological evaluation in diagnosing cochlear implant misplacement.
Journal: International Journal of Pediatric Otorhinolaryngology - Volume 75, Issue 11, November 2011, Pages 1459–1462