کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4115043 | 1606081 | 2008 | 6 صفحه PDF | دانلود رایگان |

SummaryObjectivesMagnet displacement is recognized as an uncommon, major complication following cochlear implant surgery. This complication necessitates operative intervention exposing patients to further surgical risk and morbidity. The purpose of this paper is to describe a unique surgical technique for securing internal magnets which have dislodged from their silicone housing within the internal receiver-stimulator.MethodsA case involving a 5-year old cochlear implant patient is reviewed. This patient experienced magnet displacement on two separate occasions. A new lasso technique for re-securing the internal implant magnet is described in detail.ResultsThe patient's implant remains fully functional with a securely placed internal magnet 36 months post-operatively.ConclusionsWhen compared to other surgical options, the “lasso” technique represents a simple, safe, efficient and cost-effective surgical option with minimal morbidity.
Journal: International Journal of Pediatric Otorhinolaryngology - Volume 72, Issue 7, July 2008, Pages 1071–1076