BackgroundThere has been concern over the effect of vigorous bubbling on the delivery pressure during the operation of the bubble nasal continuous positive airway pressure (CPAP) system. We investigated the relationship between intra-tubing pressure changes and flow rates in a closed bubble CPAP system in vitro.MethodsUsing an experimental (in vitro) model, the distal connecting tube of the CPAP system was immersed under the water seal to a depth of 5 cm. Sixteen different flow rates, ranging from 2 L/min to 20 L/min, were tested. The procedure was repeated 10 times at each flow rate, and the intra-tubing pressure was recorded.ResultsThe intra-tubing pressure within the model increased as the air flow rates were adjusted from 2 L/min to 20 L/min. The relationship was represented by the following equation, pressure (cmH2O) = 5.37 + 0.15 × flow rate (L/min) (R2 = 0.826, p < 0.001).ConclusionThese results demonstrated that the intra-tubing pressure in a bubble CPAP system was highly correlated with flow rate in vitro.
Journal: Pediatrics & Neonatology - Volume 51, Issue 4, August 2010, Pages 214-218