کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4221049 1609595 2016 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Wide-detector axial CT versus 4 cm detector helical CT for transcatheter aortic valve replacement: iodine dose, radiation, and image quality
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
Wide-detector axial CT versus 4 cm detector helical CT for transcatheter aortic valve replacement: iodine dose, radiation, and image quality
چکیده انگلیسی

PurposeThis study aims to compare transcatheter aortic valve replacement (TAVR) planning on 16 cm wide-detector computed tomography (CT) to TAVR planning on 4 cm detector CT.Materials and methodsA total of 36 patients who had TAVR planning axial CT on a wide-detector scanner (protocol 1) were compared to 36 patients who had helical 4 cm detector CT (protocol 2).ResultsVascular attenuation was greater for protocol 1, but image noise, contrast-to-noise ratio, and signal-to-noise ratio were the same. Radiation dose was lower and iodine dose was less for protocol 1.ConclusionProtocol 1 had greater vascular attenuation and similar image quality but lower radiation and less iodine compared to protocol 2.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Imaging - Volume 40, Issue 6, November–December 2016, Pages 1213–1218
نویسندگان
, , , , , , ,