کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5031923 1471103 2017 29 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental and computational analysis of micromotions of an uncemented femoral knee implant using elastic and plastic bone material models
ترجمه فارسی عنوان
تجزیه و تحلیل تجربی و محاسباتی از میکروموتورهای ایمپلنت زانو بدون استخوان فمورال با استفاده از مدل مواد استخوانی الاستیک و پلاستیک
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
چکیده انگلیسی
TKR surgical cuts were applied to five cadaveric femora and micro- and clinical CT- scans of these un-implanted specimens were made to extract geometrical and material properties, respectively. Micromotions at the interface were measured using digital image correlation. Cadaver-specific FEA models were created based on the experimental set-up. The average experimental micromotion of all specimens was 53.1 ± 42.3 µm (mean ± standard deviation (SD)), which was significantly higher than the micromotions predicted by both models, using either the plastic or elastic material model (26.5 ± 23.9 µm and 10.1 ± 10.1 µm, respectively; p-value < 0.001 for both material models). The difference between the two material models was also significant (p-value < 0.001). The predicted damage had a magnitude and distribution which was comparable to the experimental bone damage. We conclude that, although the plastic model could not fully predict the micro motions, it is more suitable for pre-clinical assessment of a press-fit TKR implant than using an elastic bone model.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Biomechanics - Volume 61, 16 August 2017, Pages 137-143