کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
5131680 | 1491317 | 2017 | 9 صفحه PDF | دانلود رایگان |
Newborn screening for diagnosis of phenylketonuria, homocystinuria, and maple syrup urine disease have been conducted by analyzing the concentration of target amino acids using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-ToF MS) based on parylene-matrix chip. Parylene-matrix chip was applied to MALDI-ToF MS analysis reducing the matrix peaks significantly at low mass-to-charge ratio range (m/z < 500). Reproducibility of inter-spot and intra-spot analyses of amino acids was less than 10%. Methanol extraction was adopted for simple and rapid sample preparation of serum before mass spectrometric analysis showing 13.3 to 45% of extraction efficiency. Calibration curves for diagnosis of neonatal metabolic disorders were obtained by analyzing methanol-extracted serum spiked with target amino acids using MALDI-ToF MS. They showed good linearity (R2 > 0.98) and the LODs were ranging from 9.0 to 22.9 μg/mL. Effect of proteins in serum was estimated by comparing MALDI-ToF mass spectra of amino acids-spiked serum before and after the methanol extraction. Interference of other amino acids on analysis of target analyte was determined to be insignificant. From these results, MALDI-ToF MS based on parylene-matrix chip could be applicable to medical diagnosis of neonatal metabolic disorders.
Journal: Analytical Biochemistry - Volume 530, 1 August 2017, Pages 31-39