کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5354133 1503583 2016 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Surface analysis and electrochemistry of a robust carbon-nanofiber-based electrode platform H2O2 sensor
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی تئوریک و عملی
پیش نمایش صفحه اول مقاله
Surface analysis and electrochemistry of a robust carbon-nanofiber-based electrode platform H2O2 sensor
چکیده انگلیسی
A vertically aligned carbon nanofiber-based (VACNF) electrode platform was developed for an enzymeless hydrogen peroxide sensor. Vertical nanofibers have heights on the order of 2-3 μm, and diameters that vary from 50 to 100 nm as seen by atomic force microscopy. The VACNF was grown as individual, vertically, and freestanding structures using plasma-enhanced chemical vapor deposition. The electrochemical sensor, for the hydrogen peroxide measurement in solution, showed stability and reproducibility in five consecutive calibration curves with different hydrogen peroxide concentrations over a period of 3 days. The detection limit was 66 μM. The sensitivity for hydrogen peroxide electrochemical detection was 0.0906 mA cm−2 mM−1, respectively. The sensor was also used for the measurement of hydrogen peroxide as the by-product of the reaction of cholesterol with cholesterol oxidase as a biosensor application. The sensor exhibits linear behavior in the range of 50 μM-1 mM in cholesterol concentrations. The surface analysis and electrochemistry characterization is presented.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Surface Science - Volume 384, 30 October 2016, Pages 251-257
نویسندگان
, , , , ,