Interventional devices with ferromagnetic components can be manipulated remotely using forces induced by the MRI gradients. To deflect the tip of an endovascular catheter, large ferromagnetic spheres of 2Â mm diameter are required to exert sufficiently high magnetic forces; however, tracking of these devices is difficult due to the large image artifacts. In this study, a new dual-echo technique is proposed to improve the stability of localizing and tracking medical devices with ferromagnetic components. MR tracking methods with selective off-resonant excitation and phase compensation with a rephasing gradient can detect ferromagnetic spheres up to a diameter of 1Â mm only. In this work, a dual-echo technique is used with two rephasing gradients to stabilize the off-set localization. With rephasing being applied in orthogonal directions, an SNR of 5 was achieved in the signal projections. Compared to a single-echo acquisition the dual-echo method reduces the position error in a phantom from 8Â mm to 1.6Â mm. In an in vivo study a tracking precision of 4Â mm was measured without steering gradients at an image update rate of 2 images per second. Steering experiments were successfully performed with a prototype catheter with ferromagnetic sphere in an aorta phantom and in the vena cava of a pig.
Journal: Journal of Magnetic Resonance - Volume 234, September 2013, Pages 176-183