کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5407092 1393202 2008 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
High-flip-angle slice-selective parallel RF transmission with 8 channels at 7 T
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی تئوریک و عملی
پیش نمایش صفحه اول مقاله
High-flip-angle slice-selective parallel RF transmission with 8 channels at 7 T
چکیده انگلیسی
At high magnetic field, B1+ non-uniformity causes undesired inhomogeneity in SNR and image contrast. Parallel RF transmission using tailored 3D k-space trajectory design has been shown to correct for this problem and produce highly uniform in-plane magnetization with good slice selection profile within a relatively short excitation duration. However, at large flip angles the excitation k-space based design method fails. Consequently, several large-flip-angle parallel transmission designs have recently been suggested. In this work, we propose and demonstrate a large-flip-angle parallel excitation design for 90° and 180° spin-echo slice-selective excitations that mitigate severe B1+ inhomogeneity. The method was validated on an 8-channel transmit array at 7 T using a water phantom with B1+ inhomogeneity similar to that seen in human brain in vivo. Slice-selective excitations with parallel RF systems offer means to implement conventional high-flip excitation sequences without a severe pulse-duration penalty, even at very high B0 field strengths where large B1+ inhomogeneity is present.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Magnetic Resonance - Volume 195, Issue 1, November 2008, Pages 76-84
نویسندگان
, , , , , , , , ,