کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5407455 1393217 2007 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A direct modulated optical link for MRI RF receive coil interconnection
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی تئوریک و عملی
پیش نمایش صفحه اول مقاله
A direct modulated optical link for MRI RF receive coil interconnection
چکیده انگلیسی
Optical glass fiber is a promising alternative to traditional coaxial cables for MRI RF receive coil interconnection to avoid any crosstalk and electromagnetic interference between multiple channels. A direct modulated optical link is proposed for MRI coil interconnection in this paper. The link performances of power gain, frequency response and dynamic range are measured. Phantom and in vivo human head images have been demonstrated by the connection of this direct modulated optical link to a head coil on a 0.3 T MRI scanner for the first time. Comparable image qualities to coaxial cable link verify the feasibility of using the optical link for imaging with minor modification on the existing scanners. This optical link could also be easily extended for multi-channel array interconnections at high field of 1.5 T.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Magnetic Resonance - Volume 189, Issue 1, November 2007, Pages 130-138