کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5428646 1508685 2013 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Hot gas flow cell for optical measurements on reactive gases
ترجمه فارسی عنوان
سلول جریان گاز داغ برای اندازه گیری های نوری بر روی گازهای واکنش پذیر
موضوعات مرتبط
مهندسی و علوم پایه شیمی طیف سنجی
چکیده انگلیسی


- New design presented for a gas flow cell for reactive gases up to 800 K.
- Design avoids contact with optical windows.
- Design verified by different methods for the IR and UV regions.
- High-resolution absorption cross-section presented for SO2 in UV up to 773 K.

A new design is presented for a gas flow cell for reactive gases at high temperatures. The design features three heated sections that are separated by flow windows. This design avoids the contact of reactive gases with the material of the exchangeable optical windows. A gas cell with this design was validated for high resolution measurements at temperatures of up to 800 K (527 °C) in the ultraviolet (UV) and infrared (IR) regions (190-20 000 nm). Verification of the gas temperature in the cell is provided by a thermocouple and emission/transmission measurements in the IR and UV regions. High-resolution measurements are presented for the absorption cross-section of sulfur dioxide (SO2) in the UV range up to 773 K (500 °C).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Quantitative Spectroscopy and Radiative Transfer - Volume 130, November 2013, Pages 392-399