کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
5529999 | 1401711 | 2017 | 6 صفحه PDF | دانلود رایگان |

Background and purposeTo analyze treatment plan robustness and plan optimization strategies of 106Ru eye plaque brachytherapy using a novel software tool.Materials and methodsA treatment planning software was developed that allows to calculate dose-volume metrics. Plaque misplacements were simulated and evaluated with respect to the effect on tumor coverage and dose changes in critical structures. Two treatment plan optimization approaches were analyzed: (a) reducing plaque size and (b) shifting the plaque away from organs-at-risk (OAR).ResultsMaximum tumor sizes were identified which can be covered by the prescribed dose for different robustness levels (0-2 mm). For an apex height of 5 mm a 1 mm uncertainty yielded changes in D2% to the lens of up to ±13 Gy in anterior and ±20 Gy to the optic nerve in posterior tumors. By reducing the plaque size Dmean and D2% to lens, optic nerve and macula were decreased by >60% for most simulated cases. Similarly, by shifting the plaque away from the lens dose reductions of 15%/mm in anterior and even 30%/mm in central tumors were achieved.ConclusionCritical structures in the treatment of uveal melanomas with 106Ru plaques can benefit from the proposed, computational treatment plan optimization.
Journal: Radiotherapy and Oncology - Volume 123, Issue 1, April 2017, Pages 119-124