Adaptive bolus chasing computed tomography angiography: Control scheme and experimental results
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر پردازش سیگنال
پیش نمایش صفحه اول مقاله
Adaptive bolus chasing computed tomography angiography: Control scheme and experimental results
چکیده انگلیسی

In this paper, a new adaptive bolus-chasing control scheme is proposed to synchronize the bolus peak in a patient's vascular system and the imaging aperture of a computed tomography (CT) scanner. The proposed control scheme is theoretically evaluated and experimentally tested on a modified Siemens SOMATOM Volume Zoom CT scanner. The first set of experimental results is reported on bolus-chasing CT angiography using realistic bolus dynamics, real-time CT imaging and adaptive table control with physical vasculature phantoms. The data demonstrate that the proposed control approach tracks the bolus propagation well, and clearly outperforms the constant-speed scheme that is the current clinical standard.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biomedical Signal Processing and Control - Volume 3, Issue 4, October 2008, Pages 319–326