ProcedureDynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
پیش نمایش صفحه اول مقاله
ProcedureDynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems
چکیده انگلیسی

PurposeWe aimed to compare the effects of a blood pressure transducer system specifically manufactured to limit underdamping artifacts with those of a standard system on hemodynamic parameter estimation and accuracy.Materials and MethodsForty-three consecutive patients undergoing vascular surgery at the University of Florence, Italy, were included. Arterial blood pressure signal was simultaneously registered with 2 MostCare monitors, connected to the artery either by a standard transducer or a specific transducer manufactured to avoid underdamping artifacts (Resonance Over-Shoot Eliminator [R.O.S.E.]; Becton Dickinson, Becton Drive, NJ). Patients were divided into 2 groups: absence (C group) or presence (R group) of underdamping/resonance artifacts of blood pressure signal. Systolic blood pressure, cardiac index, maximal pressure/time ratio (dP/dtMAX), and cardiac cycle efficiency were recorded every 30 seconds for 30 minutes. A total of 2675 measurements were performed with 34.9% incidence of underdamping/resonance artifacts.ResultsAll hemodynamic parameters showed clinically acceptable differences in the C group; in contrast, the results differed greatly in the R group between standard and R.O.S.E. transducer (systolic blood pressure bias, 16.7 mm Hg; cardiac index bias, 0.24 L min−1 m−2; dP/dtMAX bias, 0.92 mm Hg/ms; cardiac cycle efficiency bias, 0.018 units).ConclusionsUnderdamping/resonance artifacts frequently affect blood pressure measurement in operating rooms and intensive care units and cause severe overestimation of systolic blood pressure and incorrect estimation of hemodynamic parameters when the pulse contour method is used.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Critical Care - Volume 26, Issue 4, August 2011, Pages 415-422
نویسندگان
, , , , , , , , ,