The measurement of carboxyhemoglobin and methemoglobin using a non-invasive pulse CO-oximeter
چکیده انگلیسی

The pulse CO-oximeter (Rad-57 Masimo Corporation, Irvine, CA) allows non-invasive and instantaneous measurement of carboxyhemoglobin (COHb) and methemoglobin (MetHb) percentage level using a finger probe. However, the accuracy and reliability of the Rad-57 against the gold standard of venous or arterial blood samples have not been clearly established. Thus, the objective of this trial is to evaluate the accuracy and precision of the Rad-57 pulse CO-oximeter by comparing it with venous sampling on the same subjects. Nine healthy subjects were subjected to carbon monoxide such that it raised the COHb to 10-14% on two different days and pooled together. The COHb and MetHb were measured with a blood gas-analyzer and simultaneously with the Rad-57 as the COHb increased from 1.4 to 14%. Results were compared using linear regression and a Bland and Altman method comparison. Mean bias and precision for COHb measured with the Rad-57 was −1% and 2.5%, respectively. The mean bias and precision for MetHb measured with the Rad-57 was 0.0% and 0.3%, respectively. The ability to detect a COHb ≥10% occurred in 54% of the samples in which COHb was ≥10-14%. In conclusion, the Rad-57 provides a reading that is between −6% and +4% of the true COHb value for 95% of all samples. The Rad-57 seems to be a good substitute as a first screening test of COHb when the pulse CO-oximeter reads <15%.

► The Rad-57 provides a reading that is −6% and plus 4% of the true carboxyhemoglobin percentage for 95% of all samples when venous blood carboxyhemoglobin percentage is less than or equal to 15%. ► The ability to detect a venous blood carboxyhemoglobin percentage ≥10% occurred in 54% of the samples in which venous blood carboxyhemoglobin percentage was ≥10-14%. ► The Rad-57 seems to be a good substitute as a first screening test of COHb when the pulse CO-oximeter reads <15%. ► The Rad-57 provides a reading that is ±0.7% of the true methemoglobin percentage for 95% of all samples when the venous blood methemoglobin percentage is less than or equal to 1%.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Respiratory Physiology & Neurobiology - Volume 182, Issues 2–3, 1 July 2012, Pages 88-92
نویسندگان
, ,