Location of side branch access critically affects results in bifurcation stenting: Insights from bench modeling and computational flow simulation
ترجمه فارسی عنوان
محل دسترسی شاخه جانبی به نتایج انتقالی در استنت گذاری بفرگکتوری تاثیر می گذارد: بینش از مدل سازی نیمکت و شبیه سازی جریان محاسباتی
کلمات کلیدی
استنت، دوختن، طراحی سلول، برداشتن بوسیدن بالون، اشتباه
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
چکیده انگلیسی

BackgroundThe aim of this study was to evaluate the impact of stent design and side branch access on final strut apposition during bifurcation stenting.Methods and resultsA series of 6 different commercially available Drug Eluting Stents (DES) (n = 42) were deployed in an identical model of a coronary bifurcation. Kissing Balloon (KB) optimization was performed after either proximal or distal recrossing of the guidewire and results were analyzed by micro-Computed-Tomography.Stent design only had a minor impact on side branch lumen area free of stent struts. Similar rate of strut malapposition was observed within the bifurcation when a consistent KB optimization protocol and an optimal distal recrossing of the wire to reaccess the side branch (SB) are followed.Conversely, proximal instead of distal cell recrossing toward the side branch produced a significant lower area of the side branch lumen free of struts than an optimal distal recrossing (60.3 ± 7.1% versus 81.1 ± 8.0%, p < 0.0001), as well as a higher rate of strut malapposed toward the SB ostium (40.6 ± 6.0% versus 26.0 ± 5.7%, p = 0.0005).ConclusionsOptimal cell recrossing of the guidewire may be critical to ensure successful stent optimization in bifurcation PCI.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: International Journal of Cardiology - Volume 168, Issue 4, 9 October 2013, Pages 3623-3628
نویسندگان
, , , , , , , , ,