ObjectiveThis study applied a computer-controlled mechanical stapler to vascular end-to-end anastomosis to achieve an automatic aortic anastomosis between the aorta and an artificial graft. In this experimental study, we created a mechanical end-to-end anastomotic model and assessed the strength of the anastomotic site under high pressure.MethodsWe used a computer-controlled circular stapler named iDrive (Power Medical Interventions, Covidien plc, Dublin, Ireland) for the anastomosis between the porcine aorta and an artificial graft. Then the mechanically stapled group (group A) and the manually sutured group (group B) were compared 10 times, and we assessed the differences at several levels of pressure.ResultsTo use a mechanical stapler in vascular anastomosis, some special preparations of both the aorta and the artificial graft are necessary to narrow the open end before the procedures. To solve this problem, we established a specially designed purse-string suture for both and finally established end-to-end vascular anastomosis. The anastomosis speed of group A was statistically significantly faster than that of group B (PÂ <Â .01). The group A anastomotic sites also showed significantly more tolerance to high pressure than those of group B.ConclusionsThe computer-controlled stapling device enabled reliable anastomosis of the aorta and the artificial graft. This study showed that mechanical vascular anastomosis with the iDrive was sufficiently strong and safe relative to manual suturing.
Journal: The Journal of Thoracic and Cardiovascular Surgery - Volume 141, Issue 5, May 2011, Pages 1265-1269