Simulation and educationA manually powered mechanical resuscitation device used by a single rescuer: A randomised controlled manikin study
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Simulation and educationA manually powered mechanical resuscitation device used by a single rescuer: A randomised controlled manikin study
چکیده انگلیسی

PurposeThe goal of this randomized, open, controlled crossover manikin study was to compare the performance of “Animax”, a manually operated hand-powered mechanical resuscitation device (MRD) to standard single rescuer basic life support (BLS).MethodsFollowing training, 80 medical students performed either standard BLS or used an MRD for 12 min in random order. We compared the quality of chest compressions (effective compressions, compression depth and rate, absolute hands-off time, hand position, decompression), and of ventilation including the number of gastric inflations. An effective compression was defined as a compression performed with correct depth, hand position and decompression.ResultsThe use of the MRD resulted in a significantly higher number of effective compressions compared to standard BLS (67 ± 34 vs. 41 ± 34%, p < 0.001). In a comparison with standard BLS, the use of the MRD resulted in less absolute hands-off time (264 ± 57 vs. 79 ± 40 s, p < 0.001) and in a higher minute-volume (1.86 ± 0.7 vs. 1.62 ± 0.7 l, p = 0.020). However, ventilation volumes were below the 2005 ERC guidelines for both methods. Gastric inflations occurred only in 0 ± 0.1% with the MRD compared to 3 ± 7% during standard BLS (p < 0.001).ConclusionSingle rescuer cardio-pulmonary resuscitation with the manually operated MRD was superior to standard BLS regarding chest compressions in this simulation study. The MRD delivered a higher minute-volume but did not achieve the recommended minimal volume. Further clinical studies are needed to test the MRD's safety and efficacy in patients.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Resuscitation - Volume 82, Issue 7, July 2011, Pages 913-919
نویسندگان
, , , , , , , , ,