BackgroundDespite colonoscopy represents the conventional diagnostic tool for colorectal pathology, its undeniable discomfort reduces compliance to screening programmes.AimsTo evaluate feasibility and accuracy of a novel robotically-driven magnetic capsule for colonoscopy as compared to the traditional technique.MethodsEleven experts and eleven trainees performed complete colonoscopy by robotic magnetic capsule and by conventional colonoscope in a phantom ex vivo model (artificially clean swine bowel). Feasibility, overall accuracy to detect installed pins, procedure elapsed time and intuitiveness were measured for both techniques in both operator groups.ResultsComplete colonoscopy was feasible in all cases with both techniques. Overall 544/672 pins (80.9%) were detected by experimental capsule procedure, while 591/689 pins (85.8%) were detected within conventional colonoscopy procedure (P = ns), thus establishing non-inferiority. With the experimental capsule procedure, experts detected 74.2% of pins vs. 87.6% detected by trainees (P < 0.0001). Overall time to complete colon inspection by robotic capsule was significantly higher than by conventional colonoscopy (556 ± 188 s vs. 194 ± 158 s, respectively; P = 0.0001).ConclusionWith the limitations represented by an ex vivo setting (artificially clean swine bowel and the absence of peristalsis), colonoscopy by this novel robotically-driven capsule resulted feasible and showed adequate accuracy compared to conventional colonoscopy.
Journal: Digestive and Liver Disease - Volume 45, Issue 8, August 2013, Pages 657-662