Reliability of plantar pressure platforms
ترجمه فارسی عنوان
قابلیت اطمینان پلتفرم های کف پالت
کلمات کلیدی
فشار پایدار، قابلیت اطمینان، اندازه گیری فشار، راه رفتن،
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی

Plantar pressure measurement is common practice in many research and clinical protocols. While the accuracy of some plantar pressure measuring devices and methods for ensuring consistency in data collection on plantar pressure measuring devices have been reported, the reliability of different devices when testing the same individuals is not known. This study calculated intra-mat, intra-manufacturer, and inter-manufacturer reliability of plantar pressure parameters as well as the number of plantar pressure trials needed to reach a stable estimate of the mean for an individual. Twenty-two healthy adults completed ten walking trials across each of two Novel emed-x® and two Tekscan MatScan® plantar pressure measuring devices in a single visit. Intraclass correlation (ICC) was used to describe the agreement between values measured by different devices. All intra-platform reliability correlations were greater than 0.70. All inter-emed-x® reliability correlations were greater than 0.70. Inter-MatScan® reliability correlations were greater than 0.70 in 31 and 52 of 56 parameters when looking at a 10-trial average and a 5-trial average, respectively. Inter-manufacturer reliability including all four devices was greater than 0.70 for 52 and 56 of 56 parameters when looking at a 10-trial average and a 5-trial average, respectively. All parameters reached a value within 90% of an unbiased estimate of the mean within five trials. Overall, reliability results are encouraging for investigators and clinicians who may have plantar pressure data sets that include data collected on different devices.

► All devices had intra-platform reliability correlations >0.7 for all parameters. ► Inter-manufacturer correlations (4 plates, 2 manufacturers) were >0.7 in 52 of 56 parameters. ► All parameters reached a value <10% different from unbiased mean within 5 trials.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Gait & Posture - Volume 38, Issue 3, July 2013, Pages 544-548
نویسندگان
, , , , , ,