کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
6268168 | 1614618 | 2015 | 11 صفحه PDF | دانلود رایگان |
• CSF total protein (TP) gel and membrane staining is linear and consistent.
• TP staining is an effective loading control for CSF Western blots (WB).
• The TP loading control outperforms albumin as a CSF WB loading control.
• The TP loading control improves the sensitivity and accuracy of CSF WB.
BackgroundCerebrospinal fluid (CSF) has been used to identify biomarkers of neurological disease. CSF protein biomarkers identified by high-throughput methods, however, require further validation. While Western blotting (WB) is well-suited to this task, the lack of a validated loading control for CSF WB limits the method's accuracy.New methodWe investigated the use of total protein (TP) as a CSF WB loading control. Using iodine-based reversible membrane staining, we determined the linear range and consistency of the CSF TP signal. We then spiked green fluorescent protein (GFP) into CSF to create defined sample-to-sample differences in GFP levels that were measured by WB before and after TP loading correction. Levels of CSF complement C3 and cystatin C measured by WB with TP loading correction and ELISA in amyotrophic lateral sclerosis and healthy control CSF samples were then compared.ResultsCSF WB with the TP loading control accurately detected defined differences in GFP levels and corrected for simulated loading errors. Individual CSF sample Western blot and ELISA measurements of complement C3 and cystatin C were significantly correlated and the methods showed a comparable ability to detect between-groups differences.Comparison with existing methodCSF TP staining has a greater linear dynamic range and sample-to-sample consistency than albumin, a commonly used CSF loading control. The method accurately corrects for simulated errors in loading and improves the sensitivity of CSF WB compared to using no loading control.ConclusionsThe TP staining loading control improves the sensitivity and accuracy of CSF WB results.
Journal: Journal of Neuroscience Methods - Volume 251, 15 August 2015, Pages 72–82