Use of siliconised infant endotracheal tubes reduces work of breathing under turbulent flow
ترجمه فارسی عنوان
استفاده از لوله های اندوته ران در نوزادان سیلیکون شده، کار تنفس تحت جریان آشفته را کاهش می دهد
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی بیو مهندسی (مهندسی زیستی)
چکیده انگلیسی
When 2.5 and 3.0 mm ID ETTs were examined, laminar flow (Re <2300) across the tube was observed, and there were no clinically significant differences in the ventilation parameters between non-siliconised and siliconised tubes. Whereas, when 4 mm ID ETTs were tested, turbulent flow was observed, and PIP, Rins, Rexp and WOBvt were significantly lower (5%, 17%, 17%, and 7%, respectively) (P < 0.05), but PIF and PEF were significantly higher (8%, 14%) (P < 0.05). Thus, the silicone inner surface of ETT offers less resistance and WOBvt in presence of turbulent flow. However, artifacts observed on the surface of non-siliconised and siliconised ETTs can potentially impair ventilation.