Calibration and performance of a fast-scanning DSC—Project RHC
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی جریان سیال و فرایندهای انتقال
پیش نمایش صفحه اول مقاله
Calibration and performance of a fast-scanning DSC—Project RHC
چکیده انگلیسی

This paper presents the calibration and evaluation of a Rapid Heat-Cool DSC (RHC), a fast-scanning calorimeter capable of controlled heating rates of up to 2000 K min−1. As the use of high scan rates increases the influence of thermal lag, a thorough analysis of the instrument performance was made. This encompasses a rigorous temperature calibration procedure, a study of the influence of sample mass and heating rate on the extrapolated onset temperature, and the analysis of the symmetry of the thermal lag in heating and cooling using phase transitions in liquid crystals. The calibration of the heat flow rate is evaluated using indium for a broad range of sample masses and heating rates, and verified by heat capacity measurements on polystyrene for a wide temperature range. The performance of the RHC proves to be very good, with estimated errors on temperature and heat flow rate (heat capacity) of 0.1 K and 2%, respectively.

Figure optionsDownload as PowerPoint slideHighlights
► Calibration and evaluation of a Rapid Heat-Cool DSC (RHC).
► Fast-scanning calorimeter capable of controlled heating rates of 2000 K min−1.
► Study of influence of sample mass and heating rate on thermal lag.
► RHC thermal lag is symmetrical in heating and cooling.
► Estimated errors on temperature and heat capacity of 0.1 °C and 2%, respectively.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Thermochimica Acta - Volume 530, 20 February 2012, Pages 64–72