Designing and producing large-volume liquid gamma-ray standard sources for low radioactive pollution measurements of seawater samples by comparison between experimental and simulation results
ترجمه فارسی عنوان
طراحی و تولید مقادیر استاندارد گامای مایع با حجم بالا برای اندازه گیری میزان آلودگی رادیواکتیو نمونه های دریایی با مقایسه نتایج آزمایش و شبیه سازی
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی کنترل و سیستم های مهندسی
چکیده انگلیسی
In order to in-situ measurement of large volume water samples, using of a portable HPGe detector was considered. Because of necessity of efficiency calibration of the detector for the geometry (100 L), the large volume standard sources were prepared. Before making large volume standard sources (100 L), the Monte Carlo method has been applied in order to optimizing the calibration procedures and in agreement with experiment results, has been caused reducing the amount of produced radioactive wastes. First, the efficiency of the portable coaxial P-type HPGe detector for 1 L liquid standard sources in Marinelli beaker geometry was simulated. Then, the experimental efficiency calibration was carried out using the detector for those 1 L liquid standard sources in Marinelli beaker geometry. The detector dead layer was determined by comparison of the simulation and experimental efficiency curve results. Then, a relation between simulation and experimental measurements, that is, between pulse-height per emitted particle, F8 tally, and estimated amount of spiked radioactive solution into the 1 L distilled water in Marinelli beaker was found. Then, the efficiency calibration of the large volume liquid standard sources was simulated. The estimated amount of spiking radioactive solution into the large volume distilled water (100 L) has been taken into account dividing experimental efficiencies (in Marinelli beaker) by the simulated efficiencies (in 100 L). Finally, by spiking the large volume distilled water with the radioactive solution, efficiency calibration of the potable HPGe detector for 100 L geometry was done.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Measurement - Volume 90, August 2016, Pages 412-417
نویسندگان
, , , ,