کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
7235721 1471088 2018 16 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
OpenSim as a preliminary kinematic testing platform for the development of total knee arthroplasty implants
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
OpenSim as a preliminary kinematic testing platform for the development of total knee arthroplasty implants
چکیده انگلیسی
Our solution is to use the OpenSim simulation software to rapidly test multiple design configurations of implants. This study modeled a testing rig which characterized the motion and laxity of knee implants. Three different knee implant designs were used to test and validate the accuracy of the simulation: symmetrical, asymmetric, and anatomic. Kinematics were described as distances measured from the center of each femoral condyle to a plane intersecting the most posterior points of the tibial condyles between 0 and 135° of flexion with 15° increments. Excluding the initial flexion measurement (∼0°) results, the absolute differences between all experimental and simulation results (neutral path, anterior-posterior shear, internal-external torque) for the symmetric, asymmetric, and anatomical designs were 1.98 mm ± 1.15, 1.17 mm ± 0.89, and 1.24 mm ± 0.97, respectively. Considering all designs, the accuracy of the simulation across all tests was 1.46 mm ± 1.07. It was concluded that the results of the simulation were an acceptable representation of the testing rig and hence applicable as a design tool for new total knees.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Biomechanics - Volume 76, 25 July 2018, Pages 53-60