کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
728893 892859 2006 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Air features in spirometric transducers
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی کنترل و سیستم های مهندسی
پیش نمایش صفحه اول مقاله
Air features in spirometric transducers
چکیده انگلیسی

The report presents a metrological analysis of the influence of the expiratory air physical features on the preciseness of the expiratory parameters defined during lung ventilation. Two kinds of spirometric transducers were taken into consideration. It is theoretically analysed and experimentally confirmed that, when a volumetric transducer (a bell) is used, the air temperature inside it changes and a special correction is necessary. Moreover, its value depends on time. When a flowmeter is used (being placed near the mouth) the expired air is measured before its temperature becomes substantially lower. For this reason a correction factor should not be the same as for a volumetric transducer. Such different kinds of transducers require the usage of different temperature correction. In the opposite case the misapplicated correction made for the flowmeter can be a source of error.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Measurement - Volume 39, Issue 5, June 2006, Pages 466–476