کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
744857 1462127 2007 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A novel potentiometric membrane sensor for determination of Co2+ based on 5-amino-3-methylisothiazole
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A novel potentiometric membrane sensor for determination of Co2+ based on 5-amino-3-methylisothiazole
چکیده انگلیسی

A new poly(vinylchloride) (PVC) membrane electrode for trace level determination of Co2+ ions has been developed based on 5-amino-3-methylisothiazole as an ionophore, o-nitrophenyloctylether as a plasticizer and oleic acid (OA) as a good lipophilic additive. The electrode exhibits a Nernstian slope of 29.5 ± 0.2 mV/decade in a linear range of 1.0 × 10−1 to 6.3 × 10−7 M for Co2+ ions. The detection limit of this electrode is 3.9 × 10−7 M. It has a fast response time of 12 s and can be used for a period of 4 months without any divergence in potentials. The proposed electrode reveals a good selectivity for Co (II) over a wide variety of other tested cations and could be used in the pH range 3.3–9.0. The electrode was successfully applied as an indicator electrode for the potentiometric titration of cobalt ions with EDTA as well as for the direct determination of Co (II) in real samples.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Sensors and Actuators B: Chemical - Volume 120, Issue 2, 10 January 2007, Pages 455–461
نویسندگان
, , ,