A failure study of a locking compression plate implant
ترجمه فارسی عنوان
یک مطالعه شکست یک ایمپلنت فشرده سازی قفل کردن
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی کنترل و سیستم های مهندسی
چکیده انگلیسی


• In addition to the static loads, cyclic loading acting on the construct caused the earlier failure.
• Initiation of crack from the area of intermetallic inclusions.
• Striation spacing versus crack length was determined.

In this case study a failed locking compression plate was investigated. Such plating systems are used to provide the stability to fractured bone and fixation. The locking compression plate had been separated in two pieces. One of the fracture surfaces from the failed component was investigated for surface topographical features. The visual, optical and scanning electron microscopy results indicated the presence of beach marks, intermetallic inclusions, corrosion pits and striations indicating fatigue crack propagation and overload failure. Some corrosion damage also was documented on the fractography. This case study shows that corrosion may have initiated fatigue crack which grew by the activities of daily living causing the failure.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Case Studies in Engineering Failure Analysis - Volume 3, April 2015, Pages 68–72