Evaluation of the medical staff effective dose during boron neutron capture therapy using two high resolution voxel-based whole body phantoms
ترجمه فارسی عنوان
ارزیابی دوز موثر در کارکنان پزشکی در هنگام جذب نوترون بور با استفاده از دو فانتوم بدن با وضوح با وضوح بالا
موضوعات مرتبط
مهندسی و علوم پایه مهندسی انرژی انرژی هسته ای و مهندسی
چکیده انگلیسی
Because accelerator-based boron neutron capture therapy (BNCT) systems are planned for use in hospitals, entry into the medical room should be controlled as hospitals are generally assumed to be public and safe places. In this paper, computational investigation of the medical staff effective dose during BNCT has been performed in different situations using Monte Carlo N-Particle (MCNP4C) code and two voxel based male phantoms. The results show that the medical staff effective dose is highly dependent on the position of the medical staff. The results also show that the maximum medical staff effective dose in an emergency situation in the presence of a patient is ∼25.5 μSv/s.