Freshly excised porcine aortic heart valve cusps were subjected to a uni-axial flexural indentation test, from which the rupture characteristics and a functional stiffness parameter were determined.It was found that the flexural mechanical properties of aortic valve cusps (i) are unaffected by their coronary position and (ii) are sensitive to the effect of mechanical preconditioning.The resulting values of the cusp's flexural mechanical properties are intended as a set of reference properties which scaffolds, meant for the tissue engineering of heart valves, must approximate in order to be considered as a functional replacement.
Figure optionsDownload high-quality image (333 K)Download as PowerPoint slideHighlights
► A flexural indentation setup is developed for heart valve cusps.
► Excised aortic valve cusps are tested for their flexural properties.
► Coronary position of the cusp has no significant influence on the properties.
► The tissue is sensitive to preconditioning by previous indentation.
Journal: Journal of the Mechanical Behavior of Biomedical Materials - Volume 13, September 2012, Pages 78–84